© National Instruments Corp. 15 NI 9229E/9239E
Accuracy, NI 9229E
Measurement Conditions
Percent of
Reading
(Gain Error)
Percent of
Range
*
(Offset Error)
Calibrated max (–40 to 70 °C) ±0.13% ±0.05%
Calibrated typ (25 °C, ±5 °C) ±0.03% ±0.008%
Uncalibrated max (– 40 to 70 °C) ±1.2% ±0.55%
Uncalibrated typ (25 °C, ±5 °C) ±0.3% ±0.11%
*
Range equals 62.64 V