NI 9229E/9239E 16 ni.com
Accuracy, NI 9239E
Input noise
NI 9229E .................................... 320 μV
rms
NI 9239E .................................... 70 μV
rms
Stability
Gain drift .................................... ±5 ppm/°C
Offset drift
NI 9229E.............................. ±150 μV/°C
NI 9239E.............................. ±26 μV/°C
Measurement Conditions
Percent of
Reading
(Gain Error)
Percent of
Range
*
(Offset Error)
Calibrated max (–40 to 70 °C) ±0.13% ±0.06%
Calibrated typ (25 °C, ±5 °C) ±0.03% ±0.008%
Uncalibrated max (– 40 to 70 °C) ±1.4% ±0.70%
Uncalibrated typ (25 °C, ±5 °C) ±0.3% ±0.11%
*
Range equals 10.52 V