National Instruments 1/2-Digit User Manual Page 33

  • Download
  • Add to my manuals
  • Print
  • Page
    / 68
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 32
© National Instruments Corporation 33 NI 4070/4072 Calibration Procedure
Verifying AC Current
To verify the AC current of the NI 4070/4072, complete the following
steps:
1. Reset the calibrator.
2. Fasten the connectors on one end of the Fluke 5440 cable to the
NI 4070/4072 HI SENSE and LO banana plug connectors, and fasten
the connectors on the other end of the cable to the HI and LO calibrator
binding posts. Figure 3 shows the correct connections. Table 9 lists the
cable connections.
3. Call
niDMM_reset to reset the NI 4070/4072 to a known state.
4. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_AC_CURRENT
Range = 0.01
Resolution = 10e–9
5. Call niDMM_Read to configure the NI 4070/4072 for a current mode
before applying current.
6. Output 1 mA at 1 kHz on the calibrator with the current output set
to
NORM.
7. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 20.
8. Output 10 mA at 1 kHz on the calibrator.
9. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 20.
10. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_AC_CURRENT
Range = 0.1
Resolution = 100e–9
11. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 20.
12. Output 100 mA at 1 kHz on the calibrator.
13. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 20.
Page view 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 67 68

Comments to this Manuals

No comments